Further Capabilities; Level Indicator Field Installation Kit; Glass Draft Shield, Gd Hanger - Mettler Toledo PE360 DeltaRange Instrucciones De Manejo

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

FURTHER CAPABILITIES OF PE BALANCES: Level indicator Field Installation Kit, glass draft shield, GD hanger
There is a Level indicator Field Installation Kit available for PE balances that are frequently moved from location to location.
This allows the balance to be leveled at every new location. Thus, the accuracy of the balance remains constant.
A glass draft shield is available as an option for the PE360.
A GD hanger is built in as standard equipment in the PE balances; this is used for weighing below the balance.

Level indicator Field Installation Kit

The Field Installation Kit contains two foot screws
(28) and one level indicator (27). Installation can
be carried out either by the user himself, or by
METTLER TOLEDO Service.
Please note: The balance should be leveled before
being calibrated. This way, the leveling kit can per-
form its function.
12
Glass draft shield (for PE360)
With the PE360, it is possible to replace the plastic
draft shield that comes with the balance with the
METTLER TOLEDO glass draft shield. Use of this shield
prevents the weighing result from being affected when
using tall weighing containers.
GD hanger
For weighings below the balance, there is a GD
hanger built into the balance. Here, the object to be
weighed is hooked onto the weighing cell; this
requires an opening in the weighing table.
– Remove weighing pan and pan support; tip
balance to its side (do not place balance on
it's upper part).
– Remove cover (23).
– Hook hanger for object to be weighed onto
hook (24).
– Place balance on its feet; install pan support
and weighing pan.
– Press tare with hanger assembled; proceed
with weighing.

Hide quick links:

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Pe3600 deltarange

Tabla de contenido