Preparing and mounting the electronic
workshop torque tester
General fitting instructions
When selecting the best place for mounting the unit, take all
relevant legal provisions and safety regulations into consideration.
 The user must be able to take up a comfortable, firm stance.
 The user must have sufficient space around him to be able to
move freely.
 Lighting must be good.
 The ambient temperatures during testing must be between 
–20 °C and +60 °C.
 The electronic torque tester must be allowed to acclimatise to
ambient temperature and humidity for at least one hour before
it is used.
 The electronic torque tester must always be protected against
dirt, damp and other damaging influences.
 The application environment must be free from all vibration and
shock.