4 Pipetting
3. Align pipetting unit
The pipetting unit can turn freely in both directions.
4. Aspirating a sample
The ISO 8655 standard requires that pipette tips are pre-wetted once be-
fore the actual pipetting procedure.
a.
c.
Operating manual
50
NOTICE
a. Press the pipetting button as far as it
b.
b. Hold the device vertically and im-
c. Allow the pipetting button to steadily
Leave the tip immersed in the liquid for a
few seconds, so that the set volume is as-
pirated completely. This is particularly
important for viscous media and for
pipettes with a large volume.
Volume range
0.5 – 100 µl
100 – 300 µl
> 1000 µl
will go.
merse the tip in the liquid.
move back to its original position.
Immersion
depth
2 – 3 mm
2 – 4 mm
3 – 6 mm
Transferpette® S-8/-12 997208
Wait time
1 s
1 s
3 s