Configuring the device

In this case, repeat the fresh air adjustment. If necessary
have the sensor changed by qualified personnel.
9.5.2
Performing a 1-button calibration
All the sensors which can be calibrated are included in the
1-button calibration.
In the case of the 1-button calibration, the sensitivity of all
sensors is set to the value of the test gas. When using the
test gas cylinder 68 11 130 = mixed gas with 50 ppm CO,
15 ppm H
S, 2.5 Vol.-% CH
2
NOTICE
i
i
If a mixed gas with another composition is used, the
specified concentration values in the X-am 5000/5600
must be changed to the target values of the mixed gas
used using the Dräger CC-Vision PC software.
1. Screw the valve onto the test gas cylinder.
2. Insert the X-am 5000/5600 in the device receptacle of the
X-zone 5000 and switch on (see chapter 3.1 on page 38).
3. Press the
key and keep it pressed for 5 seconds to open
the calibration menu.
4. Enter the password (password on delivery = 001).
5. Select the 1-button calibration with the
for 1-button calibration »
6. Press the
key to start the 1-button calibration.
OK
7. Connect the hose from the test gas cylinder to the X-
zone 5000.
8. Vent the test gas into a fume cupboard or into the open air.
The currently displayed measured values start to flash. The
flashing stops after a static measured value has been
reached.
The calibration is carried out automatically.
The displayed measured values change to the values
according to the gas supplied.
When the calibration has completed and the displayed
measured values have stabilized:
The display of the current gas concentration alternates with
the display » OK «.
9. Press the
key to exit the calibration or wait for approx.
OK
5 seconds.
The X-am 5000/5600 changes to the measuring mode.
10. Disconnect the hose from the test gas cylinder from the X-
zone 5000.
If a fault has occurred during the 1-button calibration:
The fault message » X « appears and » - - « is displayed
for the respective sensor instead of the measured value.
In this case repeat the calibration.
If necessary, replace the sensor.
56
, 18 Vol.-% O
.
4
2
key. The symbol
« flashes.
10
Configuring the device
IR
To individually configure a standard-configuration device,
connect the device to a PC.
Communication is carried out via a
(order no. 83 17 409)).
The Dräger CC-Vision PC software is used to perform the
configuration.
The following settings can be configured, among others:
Horn volume
Wireless function
Alarm frequencies
Alarm pattern
Alarm forwarding
Life signal (light pattern, horn volume)
Actions of the switch relay
NOTICE
i
i
Observe the documentation and online help of Dräger
CC Vision PC software.
02833112.eps
USB DIRA III dongle
Dräger X-zone 5000