11.1 Fault messages

Special symbol »
« and
displayed numerical
code:
01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
29
30
31
32
Dräger X-zone 5000
Cause
X-am 5000/5600 with alkali supply unit.
Communication with X-zone 5000.
Communication error with battery controller
X-zone 5000.
Main battery alarm X-am 5000/5600.
Battery pre-alarm X-am 5000/5600.
Charging current X-am 5000/5600 too low.
Holder (pump) detected, but no pump fitted.
Flow fault
Holder status change (pump) during operation. Check that the holder (pump) is securely
Check sum error program code
Check sum error operating parameter
Check sum error operating parameter
Check sum error operating parameter
Working memory test error
Faulty ADC conversion.
No contact with main device in grouping mode. Check wireless connection to group device.
Charging electronics faulty.
Battery completely discharged.
Battery main alarm of the X-zone 5000.
Battery pre-alarm of the X-zone 5000.
Device error X-am 5000/5600.
Alarm pattern X-am 5000/5600 faulty.
Incorrect software version in the gas detection
instrument.
Gas type or unit unknown.
More than 8 different gases in one group.
Battery main alarm
Battery completely discharged.
Charging electronics faulty.
Communication error with battery controller
X-zone 5000.
Faults, cause, remedy
Remedy
Insert X-am 5000/5600 with battery power
pack.
Test IR interface on X-zone 5000 and on X-am
5000/5600.
Contact DrägerService.
Test charging contacts on X-zone 5000 and on
X-am 5000/5600.
Test charging contacts on X-zone 5000 and on
X-am 5000/5600.
Test charging contacts on X-zone 5000 and on
X-am 5000/5600.
Use holder for diffusion mode.
Check intake hose.
positioned.
Contact DrägerService.
Contact DrägerService.
Contact DrägerService.
Contact DrägerService.
Contact DrägerService.
Contact DrägerService.
Contact DrägerService.
Charge X-zone 5000.
Charge X-zone 5000.
Charge X-zone 5000.
Test X-am 5000/5600.
Contact DrägerService.
Have the software update performed by
DrägerService.
Test gas measuring devices individually to
ensure that all gases are known. Do not use
gas measuring devices with unknown gases.
Test sensor fitting within the group and replace
gas measuring devices if necessary or remove
from the group.
Charge X-zone 5000.
Charge X-zone 5000.
Contact DrägerService.
Contact DrägerService.
59