en
8.
Maintenance
The display "Battery replacement recommended" on the Acculan 3Ti char-
ger GA320 (see TA014535) indicates when the battery should be replaced
due to reduced capacity.
No other maintenance measures must be performed on the battery.
For corresponding services, contact your national B. Braun/Aesculap
agency, see Technical Service.
9.
Troubleshooting
►
Replace product: have defective products repaired by Aesculap Techni-
cal Service, see Technical Service.
Fault
Detection
Battery not
No signal tone when
working
inserting the battery
into the machine
Battery
Symbol "Battery
empties
replacement recom-
quickly
mended" illuminated
on the charger at the
end of the charging
process
10. Storage and transport
WARNING
Risk of injury from burns and damage to product in case of short-cir-
cuit!
►
When storing the product, ensure that the product cannot short-
circuit with other products or metal objects.
►
In order to prevent damage during transport, ship product in the orig-
inal packaging and in an outer carton.
►
If the original packaging is no longer available, ensure the following
when packaging the product:
– The battery contacts are protected against short-circuit.
– The packaging completely encloses the product.
– The product is secured against movement within the packaging.
►
Completely charge the product upon receipt and before extended peri-
ods of storage.
11. Accessories/Replacement Parts
Art. no.
Designation
TA015514
Instructions for use, battery GA666, GA676 (A4)
TA011867
Instructions for use, battery GA666, GA676 (flyer)
6
Cause
Remedy
Battery bro-
Replace prod-
ken
uct.
Battery cells
Have battery
at end of life
cells replaced by
Aesculap Tech-
nical Service.
12. Technical Service
DANGER
Danger to life of patients and users in case of malfunctions and/or
failure of protective measures!
►
Do not perform any servicing or maintenance work under any cir-
cumstances while the product is being used on a patient.
CAUTION
Modifications of medical devices may result in a loss of potential
guarantee/warranty claims and forfeiture of applicable licenses.
►
Do not modify the product.
►
Contact national B. Braun/Aesculap representative for service and
repair.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95 -1601
Fax:
+49 (7461) 14 -939
E-Mail:
ats@aesculap.de
Other service addresses can be obtained from the address indicated above.
13. Technical Data
13.1 Classification in accordance with Regulation (EU)
2017/745
Art. no.
Designation
GA666
Acculan battery
GA676
Acculan battery
13.2 Performance data, information and standards
GA666
Cell type
NiMH
DC voltage
9.6 V
Capacity
1.05 Ah
Weight
approx. 0.304 kg
Dimensions (L x W x H)
approx. 120 mm
x 43 mm x 50 mm
EMC
IEC/DIN 60601-1-2 IEC/DIN 60601-1-2
Compliance with stan-
IEC/DIN 60601-1
dards
Class
I
I
GA676
NiMH
9.6 V
1.95 Ah
approx. 0.425 kg
approx. 140 mm
x 43 mm x 50 mm
IEC/DIN 60601-1