6. Press the Start button.
The LOW CURRENT message
will turn up:
7.
Turning the Test current control
clockwise, increase current until
obtaining the desired value or
until the current bargraph reads
100 divisions. The bargraph will
show the test current value as
the percentage of the nominal
value selected by using the
Range selector.
8.
The equipment should be connected to the mains supply in order to
work with 100 A.
9.
The lowest current for measuring is 10% of the range nominal value.
It is important to consider that the measurement errors increase while
test current decreases. The error is specified for test currents higher
than 80%.
10. The Display will show the resistance value measured and the unit
used (Ω [ohms], mΩ [milli-ohms] or µΩ [micro-ohms]).
11. The value can be retained in the
second line of the display by
pressing the button Hold. If you
press this button again, the value
will be erased.