Descargar Imprimir esta página

Medela Invia Vario 18 AC/DC c/i Instrucciones De Uso página 9

Ocultar thumbs Ver también para Invia Vario 18 AC/DC c/i:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 163
Filter test
XII
Functional test
XIII
ok?
OK
1. Check to ensure the filter is working properly.
Filter open!
a. Select maximum vacuum
b. Switch on the Vario
c. Read / compare the vacuum
If the mentioned value is exceeded, the filter
must be replaced.
1. Check to ensure there is sufficient vacuum
a. Switch the device on
Before first use, completely discharge the battery
and charge for five hours
Select constant suction mode
b. Set maximum vacuum
Press knob to turn
c. Attach patient tubing to the aspirator, cover end of tubing
with your thumb. Ensure vacuum reaches -55kPa/-413mmHg.
d.
Compare vacuum
+ 6562 ft
+3280 ft
+1640 ft
not OK, Contact your service center
2. Use
Select correct vacuum and suction mode according to the
particular application
Press knob to turn. Locking safety feature will not allow
knob to turn unless it is pushed in.
MINIMUM VALUES
+ 2000 m
– 43 kPa
+ 1000 m
– 48 kPa
+ 500 m
– 51 kPa
0
– 55 kPa
LIMIT
-10 kPa
-75 mmHg
-323 mmHg
-360 mmHg
-383 mmHg
-413 mmHg
7
EN

Publicidad

loading