4 Performing a functional test
4.1 Checking the rotating unit mechanics within a test cycle
A
B
C3
9.4
P1
Prog.
S
T
R
12
4
Performing a functional test
Performing a functional test
4.1
Checking the rotating unit mechanics within a test
cycle
Checking the rotating unit mechanics within a test cycle
The test cycle is executed without radiation. The test cycle is used to
check that the unit is functioning correctly and to ensure that a complete,
uninterrupted cycle is possible. In the event of increased resistance, the
rotating unit stops automatically.
1. Use the arrow keys (A) or (B) to select a PAN program (e.g. P1).
XG5 test cycle
2. Press the T key on the control panel.
3. Press the release button.
kV
mA
4. Check the mechanics of the rotating unit during the test cycle:
5. If the cycle is interrupted or unusual noises can be heard, the
mechanics of the rotating unit are defective.
6. Inform the operator of any deficiencies.
NOTICE! The conversion may only be performed on a unit that is
intact.
7. Press the T key on the control panel again.
4.2
Checking the height adjustment mechanics
Checking the height adjustment mechanics
1. Use the Up/Down keys on the control panel to move the unit up and
down through its entire adjustment range.
2. Check the height adjustment mechanics:
3. If the height adjustment does not function correctly or unusual noises
can be heard during the adjustment (e.g. a hammering noise which
changes with the speed), the mechanics of the height adjustment or
the height adjustment motor are defective.
4. Inform the operator of any deficiencies.
NOTICE! The conversion may only be performed on a unit that is
intact.
Converting instructions for the "ORTHOPHOS XG 5 to XG 3D upgrade kit"
The program enters test cycle mode.
The test cycle is started.
Can the cycle be executed without interruption?
Are there any unusual noises during the cycle?
The program exits test cycle mode.
Does the height adjustment function correctly?
Are there any unusual noises during the adjustment?
Sirona Dental Systems GmbH
63 52 970 D3352
D3352.092.08.05.02 07.2012