147
III EMC Compliance
Note:
Warnings:
● The instrument conforms to the requirements of IEC60601- 1 - 2 , EN 60601-
1-2 and ISO 80601-2-61standards for electromagnetic compatibility.
● The user shall install and use the EMC information provided in the random file.
● Portable and mobile RF communication equipment may affect the performance
of the instrument, avoid strong electromagnetic interference when using, such
as close to the mobile phone, microwave oven, etc .
● The guidance and manufacturer's declaration are detailed in the table below.
● The instrument should not be close to or stacked with other equipment. If it must
to be close to or stacked, it should be observed and verified to be able to
operate
normally under its configuration.
● In addition to the cables sold by the instrument manufacturer as spare parts
for internal components, the use of other accessories and cables may result in
increased emission or reduced immunity.
Table 1
Guidance and manufacturer's declaration-electromagnetic emission
The Handheld Pulse Oximeter is intended for use in the electromagnetic
environment specified below. The customer or the user of the Handheld Pulse
Oximeter should assure that it is used in such an environment.
Emissions test
Conducted emis-
sions
CISPR 11
Radiated emissions
CISPR 11
Harmonic emis-
sions
IEC61000-3-2
Voltage fluctua-
tions/flicker
emissions
IEC61000-3-3
III EMC Compliance - ENGLISH
Complian-
Electromagnetic environment-guidance
ce
The Handheld Pulse Oximeter uses RF
energy only for its internal function. The-
refore, its RF emissions are very low and
Group 1
are not likely to cause any interference in
Class B
nearby electronic equipment.
The Handheld Pulse Oximeter suitable for
Class A
use in all establishments, including do-
mestic establishments and those directly
network that supplies buildings used for
domestic purposes.
Complies