Power supply
Output channel
Wave sharp
Output voltage
Output current
DC capacity of output
Treatment time
Output intensity
Treatment mode
Number of treatment
program
Operating conditions
Transportation and
storage environment:
Dimensions
Weight
Automatic shutoff
Electrode detection
function
Output precision
Grade of waterproof
Electric shock
Applied part
Mode of operation
Device service life
Note: Not intended to be sterilized.
Not for use in an OXYGEN RICH ENVIRONMENT
4.2 Technical specification for TENS mode
P.W. (pulse width)
P.R. (Frequency)
Output characteristics
Tesmed copyright - it is absolutely forbidden the partial or total reproduction of any present content
DC3.7V/400mAh Polymer lithium battery (Built-in)
Dual, isolated between output channels
Bi-phase square-wave pulse
Max. 40Vpp (at 500ohm load)
Max. 80mA (at 500ohm load)
0
10-30 minutes
0 to 40 levels, adjustable
TENS, EMS and Massage
TENS: 12 programs;
EMS: 12 programs;
Massage: 12 programs
5°C-40°C (41.0°F-104°F) with a relative humidity of 30%RH-85%RH
5°C to 40°C with a relative humidity of 15%-93%, atmospheric
pressure from 700 hPa to 1060 hPa
atmospheric 700hPa - 1060hPa
-10°C-50°C (14.0°F-122°F); 15%RH-90%RH, non-condensing
-25-55°C (-13-131°F); ≤95% RH;
700hPa – 1060hPa
110*60*16.2mm (L*W*H)
95g
3min.
The amplitude level will be reset to 0 V, when the amplitude level
is 1 or greater and an open circuit at either channel is detected.
All the output parameters allow ±20% error for the specification.
IP22
Internally powered ME equipment
Internally powered ME II equipment
Type BF applied part, including the whole unit
Continuous operation
5
2 years (Normally)
80-300us
1-150Hz (Hz=vibration per second)
Constant Voltage (CV)
110
10