application procedures outlined on electrode packing when using the new replacement
electrodes to maintain optimal stimulation and to prevent skin irritation.
Caution:
Always use the electrodes which comply with the requirements of the IEC/EN60601-1, ISO10993-
1/-5/-10 and IEC/ EN60601-1-2, such as the electrodes with CE mark, or which are legally
marketed in the US under 510(K) procedure.
5.4.2 Place electrodes on skin
Place the electrode on the part you need treatment, and base on
the instructions of this user manual. Before use, please make the
skin that need treatment and the silica gel of electrode clean and
humid, to prevent the discomfort that cause by dry skin. Please
ensure the skin and electrode connection well.
Caution:
1. Please turn off to prevent electric shock when pre-moving the treatment position of the
electrode.
2. Before applying the electrode, it is recommended to wash and degrease the skin, and then
dry it.
3. Never remove the electrode from the skin while the stimulator is still treating.
4. Only use the electrodes that provided by the manufacturer, use other accessories could
result in injury to the patient.
5. It is recommended that, at minimum 16mm
5.5 Electrode placement
FDES116 is a kind of OTC stimulator, suitable for home use, you only
Trio 6.5
need to use according to the user manual, place the electrode on
the position where you feel pain and need to exercise, treating and
adjusting base on your own feeling.
The applicable treatment part of each program was shown on LCD. You
can follow up this indication for treatment.
Please refer to te electrode treatment, program and placement
sheets from page 128 to page 160.
13
Tesmed copyright - it is absolutely forbidden the partial or total reproduction of any present content
2
self-adhering electrodes are used.