In treatment, if no further operation after 15 seconds, the operation system will be locked auto-
matically, in order to prevent unintended wrong operation, and LCD will show the lock icon. You
can press V button to unlock operation.
Caution:
If the stimulation levels are uncomfortable or become uncomfortable, reduce the stimulation
intensity to a comfortable level and contact your medical practitioner if problems persist.
6.6 Turn the device off
At the wait mode, press the [
] button for 2 sec. to turn off the device, and the LCD will be
blank.
6.7 Load detect function
When the device haven't detect the load or the electrode with skin connect not good, it will
automatically detect the load if the intensity is added to 1 level, if it is judged to no-load, the
intensity will automatically return to zero and the device will return to wait mode.
6.8 Low battery indication
When the battery voltage less than d.c. 3.3V ± 0.2V the low battery icon will be blink to showing
the LCD display. In the status, you should be connect the micro USB cables to charging.
16