An alarm will sound at the end of each phase if selected in the program. You can stop the alarm after the
last phase by pressing STOP. A run cannot be continued after pressing STOP. Disconnect the leads and
proceed with post-electrophoretic techniques. Since diffusion will begin as soon as the voltage is turned
off, you should remove the gel and begin staining, blotting or autoradiography immediately.

5.5 Optional programming

Some special functions are placed in MORE. These are:
COPY: Copying a program.
CLEAR: Clearing a program.
SETUP: Disabling the start current check.
The MORE mode is described in Fig. 8.
1. Enter MORE
2. Choose COPY or CLEAR a
program or choose SETUP for
disabling the start current check.
3. Choose Program to
copy from. (1→1)
COPY: P1 →P?
Used: P12--56-89
Selected by
Confirm with
4. Choose
Program
destination
(3→4)
COPY: P1→P3
Free:
P--34--7--
Selected by
Confirm with
5. Message
Prog. P1 copied
to P4
Wait 2s
14
SET:
500 V
Select MORE by
MORE:
copy/clear/SETUP
Selected by
Confirm with
COPY
3. Choose Program
(1→8)
CLEAR:
Used:
Selected by
set
Confirm with
enter
4. Message
Prog.
set
enter
6. Back to SET
SET:
mode
500 V
Fig. 8 . Optional programming in MORE mode.
1 1
1 Vh
1 mA
5 W
more
SETUP
set
enter
CLEAR
Program 1
P12--56789
set
enter
P8
Cleared
Wait 2s
1 1
1 Vh
1 mA
5 W
SETUP
3. Choose Program
(1→2)
SETUP:
Select program
Selected by
Confirm with
4. Choose start
current check
YES or NO
Current check at
start?
YES/no
Selected by
Confirm and
Exit by
P1
set
enter
set
enter