Introductory User Manual
Environmental Specifications
Tab. 1-3:
Environmental Specifications
Specifications
Optimal room temperature
Temperature operation range
Maximum temperature gradient
Relative humidity
Atmospheric pressure
Heat Dissipation to Air
Tab. 1-4:
Heat Dissipation to Air
The heat dissipated must be removed by a ventilation or air conditioning system.
Configuration
Heat load (with control electronics, X-ray generator
and vacuum pump)
Minimum clearance behind instrument
1-6
Data
24°C / 75°F
17 - 29°C (63 – 84°F), 'T d r 2°C (3.6°F) for very high stability of the
measuring values
2°C (3.6°F) per hour
20 % - 80 %, condensation not allowed
Spectrometry system operational at all terrestrial locations under atmos-
pheric air pressure conditions up to an altitude of 3000 meters (9800
feet).
S8-1kW
d 2.0 kW
50 cm / 20''
S8 TIGER X-ray Spectrometer
S8-3kW/4kW
d 0.5 kW
20 cm / 8''
DOC-M80-ZXX031 V4 – 02.2010