Preventative Maintenance - Stryker 1079 Manual De Uso

Camilla para cirugía ocular
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 95
English
Preventative maintenance should be performed at a minimum of annually. A preventative maintenance program should
be established for all Stryker Medical equipment. Preventative maintenance may need to be performed more frequently
based on the usage level of the product.
______
______
______
______
______
______
______
______
______
______
______
______
______
______
______
______
______
______
______
______
______
______
Product Serial Number:
Completed by: ____________________________________________ Date: _____________________
Return To Table of Contents
1-28

Preventative Maintenance

All fasteners secure
Siderails move and latch properly
Engage brake pedal and push on the stretcher to ensure that all casters lock securely
Steer function working properly
All casters secure and swivel properly
Body restraints working properly
I.V. pole intact and operating properly
Oxygen bottle holder intact and operating properly
Fowler operating and latching properly
Optional articulating head piece locking and releasing properly
Gatch operating properly
Trendelenburg/Reverse Trendelenburg operating properly
No rips or cracks in mattress cover
Arm boards intact and operating properly
Arm board support levers intact and operating properly
Ground chain intact
No leaks at hydraulic connections
Hydraulic jacks holding properly
Hydraulic drop rate set properly
Hydraulic oil level sufficient
Lubricate where required
(page
Accessories and mounting hardware in good condition and working properly
29)
1079-009-005 REV E
www.stryker.com

Publicidad

Tabla de contenido
loading

Tabla de contenido