TABLE 3
Guidance and declaration of manufacturer-electromagnetic immunity. (For home
healthcare environment)
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environment.
IMMUNITY test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Proximity fields
from RF wireless
communications
equipment IEC
61000-4-3
ART-70295-01-AA
IEC 60601 test level
3V for 0.15- 80MHz; 6V in
ISM and amateur radio
bands between
0.15- 80MHz
80% AM at 1kHz
10V/m, 80MHz -
2.7GHz, 80% AM at
1kHz
According to Table 9 of
IEC 60601-1 2:2014
Compliance level
3V for 0.15- 80MHz; 6V in
ISM and amateur radio
bands between
0.15- 80MHz
80% AM at 1kHz
10V/m, 80MHz -
2.7GHz, 80% AM at
1kHz
According to Table 9 of
IEC 60601-1 2:2014
Electromagnetic environment
guidance
Portable RF communications
equipment (including peripherals
such as antenna cables and external
antennas) should be used no closer
than 30 cm (12 inches) to any part of
the FIT Pro, including cables specified
by the manufacturer. Otherwise,
degradation of the performance of
this equipment could result.
Interference may occur in the vicinity
of equipment marked with the
following symbol:
ENGLISH | 27