Explanation of control panel indicators
dIndicator
Explanation
Segments light up in
Battery capacity depending on number of displayed segments:
red
> 75 % (4 segments)
> 50 % (3 segments)
> 25 % (2 segments)
< 25 % (1 segment)
Segments light up in
Residual capacity of particle filter
orange
segments:
> 75 % (4 segments)
> 50 % (3 segments)
> 25 % (2 segments)
< 25 % (1 segment)
Segments light up in
Flow rate intensity depending on number of displayed segments:
green
• Highest flow rate (4 segments) = 210 l/min (160 l/min with mask)
• Higher flow rate (3 segments) = 200 l/min (160 l/min with mask)
• Increased flow rate (2 segments) = 180 l/min (160 l/min with mask)
• Standard flow rate (1 segment) = 160 l/min
1)
The residual capacity of the gas filter or the gas filter components of the combination filter cannot be indicated.
2.2.2
Filters and facepieces
Filters
Always install 3 new filters of the same type! i.e. 3 x A2P R SL or 3 x ABEK2 P R SL
For P R SL filters it is possible to use only 2 new filters. (Figure B) Use filter port plug.
Install only new filters without any damage.
Caution! Before mounting new filters, make sure they are undamaged, in the original
packing and their service life has not expired (expiry date is marked directly on the body
of the filter). Check also whether the seals at the connecting points of the filters and the
blower unit are undamaged and safe.
Facepieces
Headtops and full face masks are described in separate Instructions for Use.
2.2.3
Breathing hoses
The following breathing hoses are available:
• standard hose for headtops (Part # 661370)
• hose with thread connection for full face mask (Part # 661379)
10
RSG Safety BV | Powered Air Purifying Respirator T-Air® | June 2021
depending on number of displayed
1)
GB
2.2.4
Rechargeable batteries
~ Illustration Figure C.
- Label Figure H.
The standard long life rechargeable Li-ion battery is specially designed for use with the
T-AIR PAPR blower. The rechargeable batteries reach their full capacity after 5 charge and
discharge cycles. The standard charge takes approx. 3 hours. In cases where the battery
has been completely drained charging may take up to 4 hours longer.
To prevent damage to or explosion of the rechargeable battery, charging is limited to a
temperature range of 0°C to 50°C. If this temperature range is exceeded or falls below,
charging will stop automatically and continue once the temperature range is reached
again.
2.2.5
Battery charger
Standard Charger
~ Illustration Figure C - D.
1)
Status LED
2)
Power supply
Explanation of the status LED
-
Status LED is green.
-
Status LED is red.
-
Status LED is flashing red.
When the rechargeable battery is fully charged, the charger switches automatically to
standby. In standby mode, the rechargeable battery stays fully charged at all times. In this
mode the rechargeable battery is neither overcharged nor damaged.
2.3
Functional description
The powered air purifying respirator, T-Air, is a respiratory protective device depending
on circulating air.
It filters the ambient air and makes it available as breathable air. The device continuously
takes in ambient air through the filters. The filters absorb harmful substances depending
on the filter type. In this way, the ambient air is recycled and finally reaches the
facepiece. There it is available as breathable air.
A continuous overpressure in the facepiece prevents ambient air from penetrating.
GB
RSG Safety BV | Powered Air Purifying Respirator T-Air® | June 2021
Rechargeable battery is inserted and fully
charged (standby mode)
Rechargeable battery is inserted and being
charged.
Malfunction