Do the following:
1. Leave the hazardous area.
2. Remove the facepiece (see Instructions for Use of the corresponding facepiece).
3. Switch off the blower unit by pushing the on/off button on the control panel for
approx. 3 seconds.
4. Open the belt and take off the device.
5. Clean and disinfect the device (see section 5.2 on page 17).
4.
Troubleshooting
4.1
Alarms
The type of alarm is indicated optically by red flashing of the corresponding indicator. In
addition, an acoustic alarm and the vibration alarm are triggered.
Fault
Cause
Low battery
The residual run-time of the
Charge Battery
rechargeable battery is almost
exhausted « 10 minutes).
Blocked filter
The particle filter residual
Change filter
capacity is almost exhausted «
10%).
Time expiration
Alarm notifying the user of the
Change filter
filter expiry which can be preset.
Time for service
Alarm notifying the user of time
Go to service
for service. It is necessary to
contact your supplier for regular
service.
The unit does not
Battery discharge
work at all
(check : the unit starts working
after replacing the battery with
another functional battery) Fault
of motor, electronics or charging
connector
It is impossible to
Fault of charger.
charge the battery
Bad battery connector
It is impossible to
Termination of battery life time.
charge the battery
sufficiently.
16
RSG Safety BV | Powered Air Purifying Respirator T-Air® | June 2021
Remedy
Recharge the battery
soon or replace with
fully charged battery (see
section 5.3.2 ).
Change particle or
combination filter soon
(see section 5.3.3).
Recharge the battery.
(if the problem persist,
check the battery)
Return to the producer for
repairing.
Contact your supplier.
Check the battery contact
Install a new charged
battery.
GB
5.
Maintenance and inspection
5.1
Maintenance intervals
Work to do
Before use After use
Clean and disinfect the device
Visual inspection
Check battery capacity
Replace rechargeable battery
Charge rechargeable battery
Replace filter
Check flow rate and warning
devices
5.2
Cleaning and disinfecting
CAUTION
Potential damage to components!
Only use the prescribed processes and the cleaning and disinfection agents specified
for cleaning and disinfecting. Other agents, methods, dosages and contact times may
damage the components.
Health hazard!
The undiluted agents are damaging to health if they come into direct contact with the
eyes or skin. Wear safety goggles and protective gloves when working with these agents.
5.2.1
Clean and disinfect the device
1. Dismantling the device:
a)
Separate breathing hose, facepiece and blower unit from each other.
b)
Disconnect the belt from the blower unit.
c)
If available, dismantle any accessories (e.g. hose and device sleeves).
d)
Dismantle filters (see section 5.3.4).
2. Clean the facepiece according to the appropriate Instructions for Use.
3. Cleaning the breathing hose and carrying system:
a)
Clean all parts with lukewarm water and cleaner with a soft cloth (max.
temperature: 30
C; concentration depending on the degree of contamination:
o
0.5 - 1%).
b)
Rinse all parts thoroughly under running water.
GB
RSG Safety BV | Powered Air Purifying Respirator T-Air® | June 2021
Annually
As necessary
X
X
X
X
X
X
X
X
X
17