C C A A U U T T I I O O N N
• Improper usage of the product can cause injury to the patient or operator. Operate the product only as described in this
manual.
• Do not modify the product or any components of the product. Modifying the product can cause unpredictable operation
resulting in injury to patient or operator. Modifying the product also voids its warranty.
• Do not lean on the bracket of the scale module.
• Do not use a foot end fixed IV pole on a product with the scale option.
• Always remove external equipment from the product litter when you zero the scale or weigh a patient.
• Always zero the scale at least once a week to maintain scale accuracy.
• Always use 1.5 VDC alkaline batteries.
• Always remove the batteries from the product during long periods of disuse.
• Do not load the IV pole above the safe working load of 40 lb (18 kg).
• Do not load an IV pole hook above the safe working load of 11 lb (5 kg).
• Do not use the protective top as a push/pull device. Product damage may occur.
• Do not clean, disinfect, service, or perform maintenance while the product is in use.
• Always clean V V e e l l c c r r o o ® after each use. Saturate V V e e l l c c r r o o ® with disinfectant and allow disinfectant to evaporate.
Appropriate disinfectant for nylon V V e e l l c c r r o o ® should be determined by the hospital.
• Always wipe down with clean water and dry each product after disinfecting. Some disinfectants are corrosive in nature
and may cause damage to the product. If you do not rinse and dry the product, you may leave a corrosive residue on the
surface of the product. This corrosive residue could cause premature degradation of critical components. Failure to
follow these disinfecting instructions may void your warranty.
• Do not use quaternary disinfectants formulated with glycol ethers.
• Do not use V V i i r r e e x x ® TB when you clean the crib. V V i i r r e e x x ® TB is not an approved cleaning agent for this product.
1900-009-005 Rev AB.0
3
EN