SETTING NEGATIVE PRESSURE ON THE V.A.C.® THERAPY UNITS
Activate V.A.C.® Therapy at -125mmHg continuous. Do not choose any other negative pressure
setting, intermittent or DPC modes of negative pressure.

ALARM RESOLUTIONS

KCI V.A.C.® Therapy Unit alarms should be addressed in a timely manner. Refer to the appropriate
therapy unit user manual for complete information on alarm resolutions. Refer to the
Correcting a Leak Condition section for correcting a leak in the dressing.

SPECIFICATIONS

Storage Conditions
Temperature Range .................................................................................................- 4˚F (-20˚C) to 140˚F (60˚C)
Relative Humidity Range ......................................................................................15% - 95%, non-condensing

Operating Conditions

Temperature Range .....................................................................................................41˚F (5˚C) to 104˚F (40˚C)
Atmospheric Pressure Range ............................................................................ 1060 hpa (-1253 ft /-381.9 m)
(for Optimum Performance)
Equipment not suitable for use in presence of flammable anesthetic mixture with air, oxygen or
nitrous oxide.
The PREVENA™ Therapy System is classified as a Type BF applied part under IEC 60601-1.
IP24 - Protection against solid objects greater than 12.5 mm and against liquid water sprays for
short periods of time.
All alerts are classified as low priority according to IEC 60601-1-8.
Conforms to IEC 60601-1, IEC 60601-1-6, IEC 62366, IEC 60601-1-8, IEC 60601-1-11.

CUSTOMER CONTACT INFORMATION

For questions regarding this product, supplies, maintenance or additional information about KCI
products and services, please contact KCI or a KCI authorized representative, or:
In the US call 1-800-275-4524 or visit www.acelity.com
KCI USA, Inc. 12930 IH10 West, San Antonio, TX 78249
For V.A.C.ULTA™ Therapy Units:
Select V.A.C.® Therapy only.
Do NOT select VERAFLO™ Therapy (see the previous WARNING).
See the Choose Therapy section of the V.A.C.ULTA™ User Manual
for more information.
.............................................................................
21
to 700 hpa (9878 ft / 3010 m)