Descargar Imprimir esta página

MEGABRAS MPK-203x Manual De Uso página 11

Micro-ohmímetro digital

Publicidad

Idiomas disponibles

Idiomas disponibles

4. Internal memory
This equipment can store up to 200 measured values in the internal
memory. During a measurement, press the Save button to store a value.
The equipment model and serial number, test number, date and time,
elapsed time, test current and measured value will be stored.
The internal memory is administrated by the instrument and works in a
cyclic way, this means, when the memory is full, the oldest values in the
memory will be replaced by the newest ones. To avoid lost of data,
always download the internal memory after finish the measurements.
Indicates that the measured resistance is
higher than the maximum value readable
in the selected range.
Indicates that some part of the equipment
has achieved the critical temperature.
Thus the system will cut the high current
generation.
12