11. Technical specifications
Test currents
Resistance ranges
Reading resolution
Output voltage
Maximum load resistance
Measurement principle
Continuous operation time : At 200 A this equipment may be used continuously
Thermal protection
Basic accuracy
Advanced features
Internal memory
Interface
Environmental protection
Safety class
: 100 A, 200 A (True DC).
The current may be continuously adjustable from 0
to 100 %.
: 0 up to 2000 μΩ (test current: up to 200 A).
0 up to 20 mΩ (test current: up to 100 A).
: 0.1 μΩ from 0 up to 2000 μΩ.
1 μΩ from 2 mΩ up to 20 mΩ.
: 3.5 Vdc @ 200 A.
4.5 Vdc @ 100 A.
5.8 Vdc (without load).
: 10 mΩ @ 200 A.
30 mΩ @ 100 A.
: Four-terminal, U/I.
for approx. 3 minutes before the thermal protection
activates.
At 100 A this equipment may be used continuously
for approx. 15 minutes before the thermal
protection activates.
: Protects all sensitive components, avoiding any
damage due to overheating.
: 2000 μΩ range:
± 0.5 % of reading ± 2 digits from 200.0 μΩ up to
1999.9 μΩ.
± 1 % of reading ± 2 digits from 0 up to 199.9 μΩ.
20 mΩ range:
± 0.5 % of reading ± 2 digits from 2.000 mΩ up to
20.000 mΩ.
± 2 % of reading ± 2 digits from 0 up to 1999.9 μΩ.
: Digital direct reading of very low resistances in the
alphanumerical display, with up to 4½ digits. Very
fast and accurate measurements.
: FIt allows the storage of 200 tests readings.
: USB for download the stored values.
: IP54 (with closed lid).
: Meets the requirements of IEC 61010-1.