Reacting to malfunction messages
Characteristic
Display message:
Overload
Display message:
Calibration due!
Display message:
Replace rechargeable
battery.
Display message:
Electronic fault E10
Display message:
Electronic fault E11
Display message:
Electronic fault E12
Display message:
Electronic fault E13
Display message:
Electronic fault E14
Replacing the rechargeable battery
When the battery symbol segments light up yellow,
the rechargeable battery in the TORSIOTRONIC
no longer full. However, you may continue working.
When the remaining battery symbol segments light
up red, the rechargeable battery in the
TORSIOTRONIC
is almost empty.
®
Cause
The TORSIOTRONIC
®
overloaded.
The set calibration interval has
expired.
The rechargeable battery is empty.
Electronic fault
Electronic fault
Electronic fault
Electronic fault
Electronic fault
is
®
Remedy
has been
Calibrate the TORSIOTRONIC
using suitable equipment, see
page 42.
Calibrate the TORSIOTRONIC
using suitable equipment, see
page 42.
Exchange the rechargeable battery
for a full one of the same type.
Send the TORSIOTRONIC
service department.
Send the TORSIOTRONIC
service department.
Send the TORSIOTRONIC
service department.
Send the TORSIOTRONIC
service department.
Send the TORSIOTRONIC
service department.
If the red battery symbol segments are ignored,
the TORSIOTRONIC
while.
Fastening data stored in the TORSIOTRONIC
are retained in the event that the rechargeable
battery is replaced.
If there is no rechargeable battery in the
TORSIOTRONIC
®
the time has to be synchronised with the PC
using the installed software after the
rechargeable battery has been replaced.
Place a charged rechargeable battery ready.
Release and remove the screw (1) with a Torx T8
screwdriver.
Remove the cover (2) of the rechargeable battery
compartment.
Remove the rechargeable battery (3).
®
®
to the
®
to the
®
to the
®
to the
®
to the
®
switches off after a short
®
for longer than 5 minutes,
®
41