To replace the bipolar circuit breaker, proceed as follows:
With the sterilizer off, open the side trim using the key provided with the equipment;
Access the generator's electrical control;
Remove the connected cables with the aid of a philips wrench;
With the aid of a wrench, press down on the fixing latch located at the bottom of the circuit
breaker;
Remove the circuit breaker from the control;
Reposition the new circuit breaker;
Reposition the circuit breaker input and output cables and reconnect;
Turn on the equipment and test the circuit breaker to validate its operation.
12.7 Motor circuit breaker replacement
The vacuum pump and water pump motor circuit breaker are located in the electrical control of the generator.
Disconnect the electrical power from the mains before starting the component replacement.
To replace the motor circuit breaker, proceed as follows:
With the sterilizer off, open the side trim using the key provided with the equipment;
Access the generator's electrical control;
Remove the connected cables with the aid of a philips wrench;
Remove the circuit breaker by pressing it down, relieving the fixing spring;
Reposition the new motor circuit breaker;
Reposition the input and output cables of the circuit breaker and make the connections again;
Turn on the equipment and test the circuit breaker to validate its operation.