Maintenance Manual
Steam Sterilizer
HI SPEED II
To calibrate the temperature sensors, it will be
necessary to use a calibration oven set at a reference
temperature for comparison. Proceed as below to
calibrate the temperature sensors:
Turn on the Calibration Oven and insert the
Sensor to be calibrated.
Set the Calibration Oven temperature to 134ºC
± 0,5ºC.
Check the reading of the sensor to be
calibrated with the calibration oven indication.
If there is a difference in temperature, adjust as
below.
Calculate the difference between the reading
of the Sensor to be calibrated and the
Calibration Oven.
The value found must be increased if the result
is positive or decremented, if negative, of the
standard factor of adjustment, which is 15.0.
To adjust, press on the parameter to be
modified, enter the desired value and confirm
with the "ENTER" key.
Check again the reading of the control with the
indication of the oven and if necessary adjust
again.
To exit press the "RETURN" buttons.
Temperature calibration 2.
Temperature calibration 3.
After performing the temperature calibration, press the
"X" icon to exit.
Code: MM.01.09.0003
Version / Revision: 3.001
Page: 47 of 81