Esterilizador de vapor

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 42
1
Start cycle/ test
CYCLE SELECTION
• Press the desired cycle button.
• Choose whether to change sterilisation time, drying time or cycle start date/time.
• Press the START button.
The S Fast programme does not guarantee class B sterilisation.
TEST SELECTION
We recommend performing the following tests every day before using the steriliser: Vacuum, Helix
and Bowie&Dick. Vacuum testing WARNING! Adhere to national and local guidelines regarding testing
frequency.
• Press TEST button.
• Press the desired test button.
• Press START button.
If necesary, it is possible to set a delayed start of the test.
CYCLE/TEST IN PROGRESS
The touch screen displays:
a
The current phase;
b
The remaining time of complention of the cycle;
c
Pressure in the sterilisation chamber;
d
The cycle number;
e
The sterilisation chamber temperature..
Information of cycle/test in progress.
f
PROGRAMME
STERILISATION
4'
18' 30"
20' 30"
TIME
PRE-VACUUM
3
3
3
PHASES
DRYING TIME
10'
14'
16'
Solids, porous, hollows
LOAD
wrapped and unwrapped
CYCLE
Class B
134 ÷ 137 °C
121 ÷ 124 °C
PROCESS
VALUES
2,03 ÷ 2,30 bar
1,04 ÷ 1,24 bar
MAX. LOAD
7 kg
Sterilization parameters may vary depending on the type of load, waight and conditions of the autoclave. Refer to the User Manual (chapter 5.6).
Max load trays included. * Use only the second tray from below.
END OF CYCLE/TEST
• The touch screen turns green and
the device beeps at the end of the
cycle;
• Press the "KEY" button.
Now you can open the door.
WARNING
The instruments and the
chamber are very hot.
3' 30"
3' 30''
3' 30''
/
3
2
3
/
5'
7'
4'
/
Solids,
Helix &
hollows
Unwrapped
Test
/
wrapped and
solids
B&D
unwrapped
Class S
Test
134 ÷ 137 °C
<40 °C
2,03 ÷ 2,30 bar
/
0,7 kg*
2kg
Empty chamber
2
The door is equipped with a safety device that blocks its opening in the following cases:
if the steriliser is switched off; while a cycle is running; when there is pressure in the
chamber; and when an alarm occurs. DO NOT FORCE THE DOOR OPEN. Press the 'Key'
button before opening the door. The key should change from orange to green, an open
padlock will appear.
2
Alarms
INFORMATION
NOTIFICATION
A
i
N
If the problem persists, contact the technical assistance.
DISPLAY
CAUSE
Insufficient water level in the
i20
clean water tank.
i22
The used water tank is full.
Conductivity of clean water higher than
i23
15 μS/cm, therefore too high.
The minimum pressure value
N61
has not been reached within the
set time.
A51
The expected vacuum value has
not been reached within the set time.
A70
Black-out during the cycle.
Sterilization not completed.
For more details
on the alarm you
need to press on
the relevant code.
ALARMS
SOLUTION
Fill the clean water tank.
Empty the used water tank.
Replace or fill the clean water tank with
demineralized water.
Reduce the load to make drying easier.
Run a cycle with less load. If it persists,
contact technical assistance.
Check the electrical outlet.
Dry the chamber an restart the cycle.
3