DXP DXPclave Manual Del Usuario página 5

Esterilizador de vapor

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 42
3
Maintenance
CHECK / ACTIVITY
Door gasket replacement**
Disk-door clearance check*
Movement lubrification*
Component wear check*
Filter cleaning / replacement
Bacteriological filter replacement**
Tanks cleaning*
Radiator cleaning*
* This activity can only be carried out by authorized maintainers or technicians.
** After the replacement, reset the counter in the "MAINTENANCE" menu.
COUNTER RESET
After replacing the component, enter the SETTINGS/MAINTENANCE menu and confirm to reset the
relevant counter.
the reset of the DOOR GASKET and INSPECTION counter is password protected as it is authorized
only by maintainers and technicians.
FREQUENCY
1 Year / 1.000 Cycles
1 Year / 1.000 Cycles
1 Year / 1.000 Cycles
1 Year / 1.000 Cycles
1 Year / 1.000 Cycles
6 Months / 500 Cycles
1 Year / 1.000 Cycles
1 Year / 1.000 Cycles
4
4
Safety information
WARNING
Ensure that the device is connected to a power socket with a protective earth connection.
Keep the plug in the socket until the sterilization is finished and do not use the socket for other
devices at the same time.
Do not touch the power cord with wet hands as this may cause electric shock.
Install the product with sufficient space to allow immediate removal of the electrical plug.
Turn off the power switch and disconnect the power cord before performing any maintenance
Keep explosive substances and flammable materials far from the device.
If the device overheats or emits a bad smell, turn off the power switch immediately, remove the plug
from the electrical socket and contact technical support.
Do not allow water or disinfectant liquid to enter the inside of the device as it may cause a short
circuit and electrical shock.
Avoid inadvertently touching the door or the area around the chamber while the device is in operation
or immediately after stopping the product, as these reach high temperatures and can cause burns.
Do not obstruct or cover the steam outlet on the product with other objects. In addition, avoid
inadvertently placing your face or hands near the steam outlet, as this can cause burns.
Only use NSK Dental Italy original components and spare parts.
Use tape indicators to verify whether an item was successfully sterilised.
CAUTION
Install the machine on a flat surface.
Do not sterilize liquids or objects other than medical instruments reported in the intended use.
Avoid any impact on the device. Do not drop the device.
Wash and dry objects before sterilization. Chemical detergent residues in the chamber can cause
corrosion or leave bad odours on sterilized objects.
Insert the objects to be sterilized using the racks. Directly inserting objects into the chamber may
cause sterilization problems, discolouration or even damage to the objects.
Ensure that any water has been drained before moving the device.
Sterilize the instruments in accordance with the parameters recommended by the manufacturer.
If any irregularities are noticed during use, stop the sterilization cycle and contact technical support.
It is the responsability of the user to fill the autoclave with clean demineralised water in compliant
with the EN 13060 standard, free from contaminants and other pathogens.
BACTERIOLOGICAL RISKS
If the sterilization cycle is not completed, the load, the trays and their restraint system, as well as the
inside of the chamber, should always be considered as potentially contaminated until a subsequent
sterilization cycle has been successfully completed.
The water in the recovery tank should be considered as contaminated, therefore necessary
precaution should be taken when emptying the tank. Check the integrity of the drain hose before
using it.
To avoid cross-contamination, wear a new pair of sterile gloves for each task. Take particular care to
replace the sterile gloves when loading or unloading instruments from the sterilization chamber and
during maintenance operations.
The use of contaminated water may involve residual risks. It is the responsability of the user to fill
the autoclave with clean demineralised water, free from contaminants and other pathogens. For the
water quality see chapter 5.9.3 .
5
EN