TROUBLESHOOTING
Before performing any operation, switch off the appliance and unplug the mains cable from the socket.
PROBLEM
The device is not working
No aspiration
Aspiration failure caused by
fluid spillage
Blocked float
Low and/or no vacuum power
If, after checking the conditions described above, the device still is not working properly, we recommend that you contact your
trusted dealer or an authorised FLAEM service centre nearest to you. You can find a list of all Service Centres at http://www.flaemn-
uova.it/it/info/assistenza
DISPOSAL
Aspiration unit
In accordance with Directive 2012/19/EC, the symbol on the equipment indicates that the equipment to be disposed of (excluding
accessories) is considered as waste and must therefore be subject to 'separate collection' . Therefore, the user must deliver this waste
(or have it delivered) to the separate collection centres set up by the local authorities, or hand it over to the retailer when purchasing a
new appliance of an equivalent type. Separate waste collection and subsequent treatment, recovery and disposal operations promote
the production of equipment from recycled materials and limit the negative environmental and health effects caused by improper waste
management. Unauthorised disposal of the product by the user entails the application of the administrative sanctions provided for in
the laws transposing Directive 2012/19/EC of the member state or country where the product is disposed of.
Accessories (Secretion collection vessel and piping)
Accessories to be disposed of as general waste after a sanitisation cycle.
Packaging
Product box
20
PAP
CAUSE
Power cable not correctly plugged into the
appliance socket or in the mains power
socket
Collection vessel lid not properly inserted
in the vessel
Clogged filter
Fouling on the float
Vacuum regulator all open
Protection filter blocked
Obstructed, kinked or disconnected filter
and device connection tubes
Vessel lid overflow valve closed or blocked
Dirty or obstructed or damaged pump
REMEDY
- Plug the power cable correctly into the sockets
- Insert the lid correctly into the collection vessel
- Replace the filter
- Remove the lid from the vessel, dismantle the pro-
tective device and remove the float. Then carry out
cleaning operations as per the paragraph "HYGIENIC
PREPARATION"
- Close the regulator completely and check vacuum
power
- Replace the filter
- Check the condition of the tubes, replace them if
obstructed and connect them correctly as per the
"CONNECTION DIAGRAM" in this manual
- Unlock the overflow valve, hold the device upright
- Take the device to your dealer or an authorised
FLAEM service centre