Sirona PerioScan Manual De Operador página 83

Ocultar thumbs Ver también para PerioScan:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Sirona Dental Systems GmbH
Operating Instructions PerioScan
Without
display
Malfunction: "No detection"
Operator error – Detection function was not acti-
vated
Malfunction: "No water"
Operator error – same water line was selected
a 2nd time and thus deactivated (no bar visible)
Operator error – spray ring closed
Operator error – filter clogged
Operator error – tip clogged
Operator error – wrong water line selected, bar at
another location
Operator error – external water supply shut off
Hose – spray water cartridge or hose defective
Unit – solenoid valve connector not plugged in,
solenoid valve connectors of block swapped, hoses
plugged in incorrectly, board pair defective, sole-
noid valve block defective
Unit – bottle clogged
Unit – bottle pushes itself up
Unit – hoses kinked
Unit – valve block defective, pump defective, board
pair defective, solenoid valve 2/2 defective, pres-
sure regulator defective
Malfunction: "No or incomplete tank
illumination"
Operator error – fixed water supply selected
Operator error – no tank selected
Unit – no or incomplete illumination of selected
tank
Malfunction: "No boost function"
Operating error
Operating error
Operating error
61 29 626 D 3496
D 3496.201.02.05.02
Error description
Action to be taken by the dentist
Activate detection function with key
Activate desired water line by pressing once
(bar lights up)
Turn open spray ring on handpiece
Replace filter on external water connection
Clean tip with cleaning wires or replace it if neces-
sary
Activate desired water line by pressing once
(bar lights up)
Turn on external water connection
Replace spray water cartridge or hose or inform
service engineer
Inform service engineer
Replace bottle
Replace bottle
Inform service engineer
Inform service engineer
Tank illumination must comply w. op. instructions
Tank illumination must comply w. op. instructions
Inform service engineer
Check power setting: No power boost is required
for 1 – 4 % due to the endo function
Check power setting: No further power boost is
possible with 100 %
Boost function was deactivated – reactivate it by
pressing
1
and
2
simultaneously
11 Troubleshooting
1
33

Publicidad

Tabla de contenido
loading

Tabla de contenido