Descargar Imprimir esta página

Etac Alto Manual Del Usuario página 20

Ocultar thumbs Ver también para Alto:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 54
Etac / Alto / www.etac.com
8. Head support.
9. Hand control.
10. Charging cable.
11. Control box, battery.
12. Actuators.
13. Adjustment knobs.
14. Castor.
15. Locking castor.
16. Locking, leg support.
17. Commode pan holders.
18. Emergency lowering button.
19. Pressure valve.
20. Product label.
Symbols
Symbols on the product and in this manual:
1. Warning, precaution or limitation.
2. Useful advise and tips.
3. Material for recycling.
4. The part/function is unlocked and can be moved/adjusted.
5. The part/function is locked and can not be moved/adjusted.
6. Maximum user weight (see technical data).
7. Click on the icon or scan the QR code for video.
8. Follow instructions for use.
9. Manual cleaning (see service and maintenance).
10. Machine washing (see service and maintenance).
11. Do not stand on the foot plates. Risk of injury if the device
tips over.
12. Do not sit on the front edge of the seat. Risk of injury if the
device tips over.
13. Total weight of the device including maximal user weight
14. Risk of squeezing.
Labels
Label description:
1. Unique Device Identification.
2. Product name.
3. Product description.
4. Item no.
5. Serial no.
6. CE-mark.
7. MD-mark.
8. Manufacturer.
9. *Bar code acc. to GS1-128 GTIN-14 and serial no.*
10. Type of protection.
11. Interval of use.
12. Input power.
13. Input voltage.
14. Recycle electronic equipment.
15. Class II equipement.
16. Type BF applied part (applies to the entire product).
*The device's date of manufacture can be read from the
barcode on the device. The number 11 is shown below the
barcode in brackets. The number combination after these
brackets is the date of manufacture.
Technical data
Electronic data
Operating voltage
Battery type
Typical operation time
Input voltage
Protection level against ingress of
water into the electrical parts
Volume level actuators
Electromagnetic compatibility
Guidance and manufacturer's declaration – electromagnetic emissions
See figure B
The device is intended for use in the electromagnetic environment specified
below. The user of the system must ensure that it is used in such an environ-
ment.
Emissions test
Conducted emissions
CISPR 11
Radiated emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Flicker emissions
IEC 61000-3-3
ESD Immunity
IEC 61000-4-2
Radiated RF Immunity
IEC 61000-4-3
Wireless Proximity Test
See figure C
EFT Immunity
IEC 61000-4-4
Surge Immunity
IEC 61000-4-5
Conducted RF Immunity
IEC 61000-4-6
Magnetic Fields Immunity
IEC 61000-4-8
Supply Dips Immunity
IEC 61000-4-11
Supply Interruptions Immunity IEC 61000-4-11
See Figure D
Handling
Installation upon delivery
Fully charge the battery before first use (see section
"Charging").
Hand control
Press and hold the "Stand by" button for three seconds to
activate or deactivate the hand control. The hand control is
20
25,7 V
Li-Ion
70 cycles per charge
100 – 240 V ~ , 50/60 Hz
IP46, protected against water jets
Max 50 dB
Compliance
Group 1
Class B
Group 1
The device must emit electromag-
Class B
netic energy in order to perform its
internal function. Nearby electronic
Class A
equipment may be affected.
Complies
Immunity test
Compliance/ IEC 60601 test level
± (2, 4, 6 ,8) kV
± (2, 4, 8 ,15) kV
10V/m
± 2 kV (100kHz)
± 0.5, ±1, ±2 kV
± 0.5, ±1 kV
± 0.5, ±1, ±2 kV
± 0.5, ±1 kV
± 2 kV
3Vrms
(6Vrms inside ISM / ASR bands)
30A/m
>95% dip in UT for 0.5 cycle
>95% dip in UT for 1 cycle
30% dip in UT for 25/30 cycles
Use of accessories, components or cables other than
those specified could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
Portable RF communications equipment (including pe-
ripherals such as antenna cables and external antennas)
should be used no closer than 30 cm (12 inches) to any
part of the device, including cables specified. Otherwise,
degradation of the performance of this equipment could
result.
BA22
Electromagnetic environment -
guidance
-
See Figure E
1-5