the START/STOP button. If no button is pressed, the most
recently used user memory is automatically used after
5 seconds.
Note: If the display is not full, it may be some malfunc-
tions of the display device. Try to switch on the device
again, if the display is still abnormal, please contact cus-
tomer service.
• Before the measurement, the last saved test result is
briefly displayed. If there is no measurement in the mem-
ory, the instrument always displays the value "0".
• For the first measurement the cuff is pumped up to
190 mmHg. For further measurements the pressure which
the cuff is pumped up to depends on the systolic pres-
sure of the last measurement.
Cuff air pressure is released slowly.
Note: This monitor will re-inflate automatically if the system
detects that your body requires more pressure for measure-
ment. As soon as a heart rate is detected, the heart rate
symbol
is displayed and flashes during the cuff deflating.
• When the measurement is completed the cuff will deflate
entirely. Heart rate, systolic and diastolic blood pressure
are displayed.
• You can interrupt measurement at any time by pressing
the START/STOP button
activation button to OFF.
or pushing the touch screen
• "Er" appears if it has not been possible to perform the
measurement properly. Observe the section on error mes-
sages/troubleshooting and repeat the measurement.
• The test result is saved automatically.
• To switch off and release the pressure, press the START/
STOP button
again or push the touch screen activa-
tion switch to OFF. If you forget to switch off the device,
it switches off automatically after approx. 1 minute.
Wait at least 5 minutes before taking another measurement.
7. Evaluating results
WARNING:
• This product does not and is not intended to provide a
medical diagnosis. Measurements results are for reference
only. Self-diagnosis and treatment, e.g. regarding medi-
cation, using measured results represent a risk for your
health. Always consult with a licensed physician for de-
termination of appropriate medication and dosage there-
of. Follow the instructions of your physician or licensed
healthcare provider. If you have or suspect that you have
a medical problem, promptly consult your physician. If you
have an emergency please call 911 immediately.
• Only a physician or a trained health care professional who
is familiar with your medical history is able to accurate-
ly interpret your blood pressure measurements. Consult
your physician before starting blood pressure monitoring.
19