Descargar Imprimir esta página

VBM Medizintechnik Tourniquet 2500 Manual Del Usuario página 27

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

2 TROUBLESHOOTING
Fault / failure
Indicator eye
Indicator eye does not change to green when the com-
pressed air supply is connected.
Indicator eye stays green after disconnecting from the
compressed air supply.
Manometer
Pointer does not respond.
Pointer not exactly at zero (tolerance ±10 mmHg).
Glass is damaged.
Pressure regulator
Pressure regulator cannot be turned anymore.
Pressure no longer reaches max. value (600  mmHg
±10 %).
Deflate button
Deflation with blue deflate button not possible.
Timer
Timer display is too weak or completely missing.
Leak in tourniquet system
Cause / troubleshooting
 Input pressure of the compressed air supply is missing
or is too low. Provide input pressure of min. 2 bar to
max. 10 bar.
 Indicator eye is faulty. Repair by manufacturer.
 Compressed air is only slowly released from the pres-
sure hose. A wall gas probe with one-way valve should
not be used.
 Indicator eye is faulty. Repair by manufacturer.
 Compressed air is missing (see indicator eye).
 Pointer is stuck or loose. Replace manometer (see "Ser-
vice Manual, 3 Replacement of Parts, Replacement of
manometer").
 Unit faulty. Completely replace manometer (see "Ser-
vice Manual, 3 Replacement of Parts, Replacement of
manometer").
 Improper use. Replace glass (see "Service Manual,
3 Replacement of Parts, Replacement of manometer").
 Pressure regulator overwound. Loosen pressure regula-
tor knob in accordance with instructions (see "Service
Manual, 3  Replacement of Parts, Pressure regulator
knob cannot be moved anymore").
 Pressure regulator faulty. Repair by manufacturer.
 Deflate valve is jammed or faulty. Completely replace
valve (see "Service Manual, 3  Replacement of Parts,
Replacement of deflate valve").
 Battery empty. Change battery (see "Service Manual,
3 Replacement of Parts, Replacement of battery").
 Connections not properly connected. Reconnect con-
nections.
 Valves (deflate, one-way, 3/2-way valve) are not tight.
Troubleshooting and repair by manufacturer.
- 27 / 72 -

Publicidad

loading

Este manual también es adecuado para:

15-12-50015-12-500nist