Capítulo 6. Localización de averías y pruebas
Número de error Descripción
4
Bootloader task registration has failed
5
Failed to warm up
6
Ps sensor failed or disconnected
7
Autoleak recovery activated
8
Bias Channel Pressure Ps Failed
9
Bias Channel Pressure Pt Failed
10
Bias Channel Vacuum Ps Failed
11
Bias Channel Vacuum Pt Failed
12
Invalid Current State
13
Pressure Ps valve curve has failed characterisation
14
Pressure Pt valve curve has failed characterisation
15
Vacuum Ps valve curve has failed characterisation
16
Vacuum Pt valve curve has failed characterisation
17
Invalid Curve Index
18
Invalid Curve Parameters
19
Zero Curve Volume
20
Flow decreased on entry
21
Flow at 40V smaller than linear end
22
Theoretical max flow < 40V flow
23
Curve fit error too large
24
Demand NULL
25
EXP out of range
26
LOG out of range
27
Measured NULL
28
Controller Pool
29
Control Memory
30
Invalid Next State
31
Test set must be at ground
32
Pressure Leak in Ps. Leak rate =
33
Pressure Leak in Pt. Leak rate =
34
Invalid Ps Demand
35
Ps pressure not settling
36
Ps Aim Overshoot
37
Ps aim value unobtainable
38
Invalid Ps Volume
39
Pt pressure not settling