6.2 Product Main Technical Parametric description
a) Impulse frequency:
TENS:(20-100)Hz, EMS:(1-15)Hz, FITNESS:(2-16)Hz.
b) Pulse width:
TENS: 120μs, EMS: 200μs, FITNESS: 200μs.
c) Impulse waveform: square wave
d) The individual pulsing electric quantity when in the
maximum output amplitude: >7μC.
e) The maximum output energy of individual pulse: ≤300mJ.
f) Influence of output end open circuit and short circuit: it is
able to support the influence of output end open and short
circuit, and its performance will not be allowed to weaken.
g) Adjustment of output amplitude: 0-16 levels.
h) The timer of therapeutic equipment: 15 minutes, 30 minutes,
45 minutes, =±10% franchise.
i) Safety classification: internal electric source class and type
BF equipment
j) Boundary dimension:
Main unit: 122mm * 55mm * 15.7mm;
Electrode Pads: 3 kinds
EPAD-D01: 70×52mm | EPAD-D02: 50×50mm | EPAD-D03:
50×50mm
k) Service life: the shelf life of subject device is 2 years; the use
life of the small Electrode Pad is 80 times and shelf life is 2
years.
6.3 Product Environmental Requirement
a) Normal work environmental requirements:
Environment temperature: +5℃-+40℃;
Environment humidity: 15%-93%RH;
Atmospheric environment conditions: 700hPa-1060hPa.
b) Storage environment requirements:
Environment temperature: -25℃-+70℃;
Environment humidity: 0-93%RH; Atmospheric
Environment conditions: 700hPa-1060hPa.
c) Transport environment requirements:
Environment temperature: -25℃-+70℃;
Environment humidity: 0-93%RH; Atmospheric
Environment conditions: 700hPa-1060hPa.
Note: It needs at least 30 minutes for device to return to normal
operating temperature from the minimum/maximum storage
temperature.