8. MAINTENANCE, SERVICING AND REPAIR, WARRANTY
Performance of preventive maintenance is not essential.
however, contribute to identifying potential problems before they become serious; thus,
enhancing the instrument's reliability and extending its useful operating life. Maintenance
services can be obtained from your local representative or from the manufacturer.
Defective items or equipment are to be serviced and repaired exclusively by persons
authorized by the manufacturer. All repair work shall employ original manufacturer's parts
only.
8.1 LAMP REPLACEMENT
WARNING!
CERAMIC lamps are at high internal pressure when cold and at operating temperature;
therefore, CERAMIC lamps may unexpectedly rupture resulting in discharge of hot
fragments of quartz and/or glass and metal.
Only handle lamps with protective covers in place. Do not handle lamps without their
protective covers unless government approved safety glasses, facemasks (with neck
protector, chest protector and gauntlets) are worn.
Always disconnect power cord and turn main switch off before lamp
replacement.
CAUTION
Take utmost care not to hit gas-seal edge of lamp; it may cause gas
lamp failure.
CAUTION
Do not touch the lamp and patient simultaneously when changing the
lamp.
CAUTION
Lamp power connection pins at the back of the lamp compartment may
retain stored energy. Do not contact the lamp power connectors in the back
of the lamp compartment.
CAUTION
This illuminator is designed to use Cuda
Use of any other lamp may void the warranty.
To replace the lamp:
Open the lamp compartment door.
Move the lamp positioning lever (found to the right of the lamp) clockwise, from
horizontal to vertical, and pull the lamp cartridge out. (Grasp only the horizontal top
and bottom grip flanges of the plastic housing of the lamp, pull straight out to
remove the lamp.)
Insert a new lamp, pushing firmly to assure full engagement to the power
connectors. Lower the lamp engagement lever counter-clockwise, from vertical to
®
LIT-238 CUDA
SURGICAL
(English)
Regular maintenance can,
®
Surgical p/n SYC0450 lamp module ONLY.
Date of Revision: 06/08/18
Rev. -
Page 19 of 147