NOTE
Switch position left
(delivery status)
NOTE
Switch position right
(delivery status)
!
NOTE
Switch position
The respective switch position is read after evaluating a measuring
result and after a reset.

Slide switch T1

Switch position LEFT: If the slide switch
and the instrument is switched on or the reset key pressed while the
instrument is switched on, the microcontroller executes the operating
program (firmware).
Switch position RIGHT: If the slide switch
and the instrument is switched on or the reset key pressed while the
instrument is switched on, the instrument is set to a mode which ena-
bles a firmware update via the serial interface.
If your instrument requires a firmware update, Heyl will provide you
with further detailed information.
Switch position
The switch position is only read immediately after a reset.

Slide switch T2

Use the slide switch T2
tle. The following indicator bottle sizes are possible:
Indicator
100 ml bottle
500 ml bottle
Switch position
The status of the slide switch T2 is read after resetting the indicator
display to 100% and after a reset.

Plug connector J1

The plug connector J1
portant for instrument operation.

Plug connector J2

The connection to the base circuit board is established via the plug
connector J2
using a ribbon cable.

Plug connector J3

The connection to the RS232 interface is established via the plug
connector J3
using a ribbon cable. Please refer to the section
entitled
"Serial interface
tion.
is in the left position
to determine the size of the indicator bot-
Switch position
Left
Right
is a programming interface. It is not im-
RS232" on page 26 for a detailed descrip-
Installation
is in the right position
19