Maintenance
!
ATTENTION
42
Once the measuring chamber has been fully cleaned, it can be
reinstalled. Reassemble in reverse order to disassembly.
Removal and installation of the sight glass windows
If the sight glass windows cannot be removed from the measurement
chamber, open the measurement chamber cover (4) and then care-
fully push them out from the inside. Do not apply excessive force and
do not use sharp objects that may damage the sight glass windows.
Ensure tension-free mounting of the sight glass windows. Tighten the
screws (6) equally alternating both sides. Otherwise, the sight glass
windows may break.
Once all assembly work has been completed, bleed the lines be-
fore restarting the instrument (see the section entitled
the indicator
bottle" on page 28).
1
Measuring chamber
2
Sight-glass window holder
3
Magnetic stirrer
4
Measuring chamber cover
5
Dosing needle
6
Screw M3x40
7
Screw M3x12
8
Flat seal 24x2
9
Sight-glass window 30x3
10 Hose adapter
11 Angled plug in connector
12 O-ring 3.68x1.78
13 O-ring 4.5x1.5
14 O-ring 1.78x1.78
"Inserting
(Article no 37615 (1-4bar)
or 37616 (0,3-1bar))
(Article no. 40176)
(Article no. 40050)
(Article no. 37534)
(Article no. 37536)
(Article no. 33253)
(Article no. 33246)
(Article no. 33777)
(Article no. 40170)
(Article no. 37538)
(Article no. 40157)
(Article no. 33797)
(Article no. 11264)
(Article no. 11245)