4.4

Inspection

In principle we recommend one inspection every year, and a minimum of one before usage
is resumed. All of the following checks must be performed and documented by authorized
persons:
Check the frame parts and the hinged tubes for plastic deformation, cracks and
impaired functioning.
Visually check for damage to the paintwork (danger of corrosion)
Check the operation of the wheels (free running, level rolling, axle play, tyres, profile,
condition of the rims, air pressure in the case of air tyres, floating axles, etc.)
Check the solidity and seating of all screws.
Verify the amount of grease on the metal joints of movable parts
The condition and security of the guides and axles of the steering wheels
Visually check all plastic parts for cracks and brittle spots
Check the functioning of the arm supports and footrests (locking, load, deformation,
wear and tear caused by loads, adjustment levers).
Check the functioning of the gas pressure springs and rope pulls (kinks, foreign
bodies), levers and adjusting screws. If necessary, adjust the rope pulls using the
adjusting screws.
Check the functioning of all adjusting mechanisms.
Check the operation of other detachable parts (example: head support, personal
safety belt, accessories, etc.)
Completeness of the delivery condition, instruction manual available?
The service must only be signed off in the maintenance plan if a minimum of all the above-
mentioned aspects have been checked.
4.5

Disinfection

WARNING:
   
Dangerous Products - The use of disinfectants is restricted to
authorized personnel.
WARNING:
Dangerous Products, change on irritate your skin - You should wear
   
suitable protective clothing because the disinfectants could irritate your skin. For
this purpose you should also take note of the product information of the solutions
concerned.
All parts of the wheelchair can be treated by scrubbing with a disinfectant.
All steps taken to disinfect rehabilitation equipment, their components or other accessory
parts are to be recorded in a disinfection report containing a minimum of the following
information (with product documentation appended):
Date of the
Reason
disinfection
Abbreviations used in column 2 (reason):
V = Suspected infection IF = Infection case
For a blank sheet of a disinfection book see § 9.
The recommended disinfectants for scrubbing (based on the list provided by the Robert
Koch Institute, RKI) are standing in the table below. The current state of the disinfectants
included in the RKI list can be obtained from the Robert Koch Institute (RKI) (homepage:
www.rki.de).
Specification
Table 4: Example of a disinfection book
W = Repetition I = Inspection
ƒ‰‡
Substance and
Signature
concentration
Inovys II
2011-11

Publicidad

Tabla de contenido
loading