INSTALLATION
04.1
FIRST INSTALLATION
The device must be installed in a comfortable
place suitable for its use.
The PIEZOSURGERY
white can be purchased
®
ready for it to be used instantly or may have
to be enabled by typing an activation key.
04.2
SAFETY PRECAUTIONS DURING INSTALLATION
WARNING: Contraindications.
Interference with other equipment.
Though compliant with the standard IEC
60601-1-2, the PIEZOSURGERY
nonetheless interfere with other devices
nearby. The PIEZOSURGERY
not be used near to or stacked on other
devices. However, if this were to prove
necessary, you must check and monitor
correct operation of the device in that
WARNING: Contraindications.
Interference from other equipment.
An electrical scalpel or other electro-
surgical units near the PIEZOSURGERY
white device may interfere with its correct
operation.
CAUTION: The electrical system of the
premises in which the device is installed
and used must be compliant to the norms
in force and to the relative electrical safety
precautions.
CAUTION: To avoid any risk of electric
shock this device must be grounded.
WARNING: Do not install the
device in places where there is a risk
of explosion. The device cannot operate
in environments where anesthetic or
In the case that your device requires an
activation key, the procedures to follow may
vary from country to country.
Always refer to your retailer to obtain
information on the above.
in way so that the power switch is easily
reachable, since this switch is considered
white may
as a load-break switch.
®
white must
®
peristaltic pump is open, the foot pedal
of the PIEZOSURGERY
be operated. The parts in movement may
cause harm to the operator.
place protected against collisions or
against accidental sprays of water or
liquids.
®
above or near heat sources. Foresee
adequate air circulation around the device
when installing it. Leave adequate space,
especially near the fan placed on the back
part of the device.
to direct sunlight or to sources of UV light.
transported, but it must be handled with
care when it is displaced. Position the foot
pedal on ground, so that it can only be
activated intentionally by the operator.
handpiece cord to the device, make sure
that the electrical contacts are perfectly dry.
If need be, dry them with compressed air.
04
INSTALLATION
CAUTION: Always position the device
WARNING: When the lid of the
white must not
®
WARNING: Install the device in a
WARNING: Do not install the device
CAUTION: Do not expose the device
CAUTION: The device can be
CAUTION: Before connecting the
EN
11