Guidance and manufacturer's declaration - Electromagnetic immunity
The PIEZOSURGERY
white is intended for use in the electromagnetic environment
®
The customer or user of the PIEZOSURGERY
environment.
Immunity
test
Conducted RF
3 Veff
IEC 61000-4-6
from 150 kHz
to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
from 80 MHz
to 2,5 GHz
N.B.:
(1) at 80 MHz and 800 MHz, the higher frequency range applies.
(2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
PIEZOSURGERY
white is used exceeds the applicable RF compliance level given above, the PIEZOSURGERY
®
white should be observed to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as re-orienting or relocating the PIEZOSURGERY
®
Test level
Compliance
IEC 60601
level
The device
continues to
work regularly
and in safety
08
TECHNICAL DATA
white should ensure that it is used in such an
Electromagnetic environment
Portable and mobile RF
communications equipment should
be used no closer to any part of the
device including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in metres (m).
transmitters, as determined by an
electromagnetic site survey
be less than the compliance level in
each frequency range
Interference may occur in the vicinity
of equipment marked with the
following symbol:
white.
®
Guidance
, may
a
.
b
®
EN
27