7 Malfunctions
7.1 Error messages
Error
Suction air not present; suction
device works when on/off switch is in
the On position.
Software error message:
Suction device vacuum switch
22
7

Malfunctions

7.1

Error messages

Possible cause
Suction tube blockage
Leak on suction tube or socket
Fault inside the processing machine
Suction power controller is at min on
the control panel.
Filter bag / HEPA filter occupied
Suction tube blocked
Interface cable defective
Fault inside the processing machine
Operating Instructions Suction devices for inLab MC X5 and
Remedy:
Check suction tube for open passage;
remove blockages.
Check suction tube for leak tightness
and tight connection to the suction
unit and processing machine; seal the
tube / connect it tightly if required.
Contact services
Set suction power controller to max.
Replace filter bag
Replace HEPA filter
Check suction tube for open passage;
remove blockages if necessary.
Replace interface cable
Contact services
Dentsply Sirona
64 98 609 D3608
D3608.201.02.04.02 09.2017