8.
Sterilization
- Steam sterilisation
134 °C / 273°F during
9.
Storage
- Keep devices
B. Filling material
Operation
Operating mode
1.
Disinfection
- Immerse the
6.8) REcOMMENDED SPEEDS
Diamond Round Bur Iso 016:
Transmetal Bur:
Carbide Round Bur Iso 010:
Carbide Round Bur Iso 014:
Tapered Diamond Bur Iso 016:
X-Gates:
Endo-Z:
DFU CAVITY ACCESS KIT - F190342.X/ 06 / 2005 updated 05/2010
- The instruments, posts and the plastic supports must be
at:
sterilized according to the packaging labelling.
- Use only autoclaves that are matching the requirements of EN
18 min.
13060, EN 285.
- Use a validated sterilisation procedure according ISO 17665
- Respect the maintenance procedure of the autoclave device
given by the manufacturer.
- Use only this recommended sterilization procedure.
- Control the efficiency (packaging integrity, no humidity, colour
change of sterilisation indicators, physico-chemical integrators,
digital records of cycles parameters).
- Traceability of procedure records
- Sterility cannot be guaranteed if packaging is open, damaged
in sterilization
or wet.
packaging in a
- Check the packaging and the medical devices before using
dry and clean
them (packaging integrity, no humidity and validity period).
environment
Warning
- Do not use disinfecting solutions containing Phenol or any products which are
obturation devices
not compatible with the treated filling material. (See general recommendation).
in NaOCl (2,5%
at least) during
5 mn at ambient
temperature.
120 000
to
160 000 rpm
150 000
to
200 000 rpm
250 000
to
300 000 rpm
150 000
to
200 000 rpm
120 000
to
160 000 rpm
500
to
160 000
to
300 000 rpm
DENTSPLY MAILLEFER
1338 Ballaigues
Switzerland
X X X X
X X X X
800 rpm
6 / 72