Problem
Probable Cause
Handpiece
Tip oscillates, but
LED does not
Handpiece LED
illuminate.
turns on and off
(Varios2 Lux)
Loss of the power
Power output is
output without
set 8 at G
operation
* Repairs cannot be made by the customer.
13. Spare Parts
Model
Water Tube Set
Water Supply
Connector
Water Filter
Spanner Wrench
(5x8) X 2 pcs
Tip Wrench
(CR-10)
Autoclavable up to max 135°C.
14. Disposing product
Consult with dealer from whom you purchased it about waste disposal.
15. Warranty
Manufacturer warrants its products to the original purchaser against defects in material and workmanship under normal
practices of installation, use and servicing. Such expendable items as O-Rings are not covered by this warranty.
Cause
The handpiece is not connected into
the Handpiece Cord correctly
Safety function is activated
Products
Order code
U1007080
U387030
U387042
Y1001301
Z221076
Firmly insert the handpiece into the Handpiece
Cord inmost
Powerful output will weaken automatically while
continuous operation is over 10min at the setting
of Maximum power at G mode. Releasing the foot
from the Foot Control. Decrease the Power less
than 5, once then increase the power again. ( Refer
to 7-5 )
Model
Tip Holder
Tip Cover S
O-Ring
Double-Face Tape
( For Control Unit Holder )
Double-Face Tape
( For Handpiece Holder )
14
Solution
Products
Order code
Z221080
Z217851
0310020080
20002545
20002544