Promax PROLITE-60 Manual Del Usario página 86

Analizador de espectros ópticos
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Measurement parameters
For each one WDM carrier (Wavelength Division Multiplexing):
λ
Wavelength where is it located (nm)
PWR
Power level (dBm).
SNR
Optical signal-noise ratio measurement (SNR).
GAIN
Difference at each measured
SETUP Menu Access to measurement parameters configuration menu.
Options
Start Lambda (nm)
Stop Lambda (nm)
Peak Excursion (dB)
Peak Threshold (dBm)
Noise Level (nm)
Close Menu
File
Options
Save Values
Manage Files
Select
Delete
Load
Close Menu
Exit
Page 32
λ
between input and output.
Sets the start wavelength for the scan.
Sets the stop wavelength for the scan.
Sets the peak excursion value. This is the amount of
amplitude that trace must raise and fall to be
considered as a peak. Lower values lead to more
signals being discerned, but if the peak excursion is set
too low, peaks in the noise floor may be discerned as
signals.
Sets the peak threshold value. Power levels below this
threshold are not considered for peak search.
Selects the offset value to measure the noise floor. This
value is used in the OSNR calculation.
Quits measurement parameters configuration menu
Data file menu access.
Saves in a file the registered measurement parameters
corresponding to the visualised trace.
Shows the data file list saved in memory.
Selects a data file containing the measurement
parameters previously described.
It allows deleting a data file.
Load a data file.
Quits data file menu.
Quits measurement function.
USER'S MANUAL. PROLITE-60
09/2007

Publicidad

Tabla de contenido
loading

Tabla de contenido