Mettler Toledo IND780 Manual De Instalación página 44

Ocultar thumbs Ver también para IND780:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

IND780 Installation Manual
• A metal cable gland is provided with the Analog Load Cell and POWERCELL
options. To further protect the IND780 from external influences, the cable's
shield wire can be spread out and pressed into cable gland by the grommet, as
shown in Figure 2-24.
• Press the grommet into the body of the cable grip, as shown in Figure 2-25.
• Move the cable through the grommet to adjust its length within the enclosure.
When making cable terminations inside the harsh enclosure, ensure that the
cable length from the terminal strip/connector to the terminal housing is
sufficient so that no strain is placed on the connector assembly when the
housing is in the fully open position.
• Finally, tighten the nut onto the body of the cable grip. Figure 2-26 shows the
assembled cable grip.
2-16
Figure 2-24: Installing Metal Cable Gland with Shield Wire
Grommet in
Cable Grip
Figure 2-25: Grommet in Body of Cable Grip
Figure 2-26: Cable Grip Assembled

Publicidad

Tabla de contenido
loading