Master Reset Button - Mettler Toledo IND780 Manual De Instalación

Ocultar thumbs Ver también para IND780:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

IND780 Installation Manual
Switch
S-1
S-2
Note that the switches shown in Figure 2-77 are both in their OFF position.

Master Reset Button

To erase all programming in the terminal and reset all settings (except
metrologically significant scale calibration data) to their factory default values,
press the master reset button adjacent to the battery. Figure 2-78 indicates the
location of this button. This process is described in the IND780 Technical Manual,
Chapter 4.0, Service and Maintenance.
To reset scale data as well, S2 (shown in its OFF position in Figure 2-78)
must be set to ON before the master reset.
2-44
S2
S1
Figure 2-77: Main PCB Switch Locations
Table 2-8: Main PCB Switch Functions
Metrology Security Switch.
When in the ON (right) position, this switch prohibits access to Scale
settings in the menu and other metrologically significant areas.
Task Expert Disable Switch.
When in the ON (down) position, disables the task expert auto-start
feature. This is used primarily for maintenance purposes.
Function

Publicidad

Tabla de contenido
loading

Tabla de contenido