PROBLEM
The silicone gasket on the door and/or the edge of the chamber where the
There is a
gasket rests are dirty.
microleak in
the door.
The silicone gasket on the door is damaged and/or out of place.
There is an object between the door and the sterilisation chamber.
The door does
The system that detects the door is not working.
not close.
The closing mechanism is broken.
The warning
The buzzer is not working.
signal cannot
be heard.
The alarm level is set low.
The date/time
on the autoclave
The internal battery of the main card has run out.
is wrong.
Excessive
The bags are not placed correctly.
condensation on
the instruments.
Excessive amount of materials to sterilise in the chamber.
The wrong sterilisation temperature has been selected.
The instruments
come out black.
Inappropriate Material.
Top distilled water level detector is not working properly.
The water tank
The connectors of the level detector of distilled water are not connected.
does not fill.
The quick connection union joints have not been placed properly or there is no
distilled water.
Distilled water has not been used.
There is dirt on the instruments.
The instruments
come out rusty
The instruments have been in contact with other materials.
or stained.
One of the instruments was rusty or about to rust.
There are white
marks (limestone/
mineral
Distilled water has not been used.
sediments) in the
sterilisation
chamber.
The printer has run out of paper.
The ink ribbon has run out.
The paper is stuck.
The printer is
not printing.
The printer is not connected to the power supply or it is switched off.
The number of copies selected is 00 on the machine.
The printer does not work.
The waste water
The internal tube is blocked/obstructed.
tank is
not emptying
The union joint that empties the waste water is obstructed or
properly.
not connected properly.
The date/time
on the autoclave
The internal battery of the main plate has run out.
changes after a
period of time.
The countdown
The internal clock/calendar does not work properly.
does not start.
The filling and/or
emptying union
There is dirt/particle on the union joint.
joints are leaking.
20
CAUSE
SOLUTIONS
Clean the silicone gasket on the door, the groove and the edge of the chamber
where the gasket rests.
Replace the silicone gasket on the door.
Remove the object.
Contact the Technical Support Dept.
Contact the Technical Support Dept.
Contact the Technical Support Dept.
Set the alarm level on the options menu.
Contact the Technical Support Dept.
See Annex 9.1 'Hygiene/sterilisation process'.
See Annex 9.1 'Hygiene/sterilisation process'.
Adjust the sterilisation temperature according to the instructions provided by
the material manufacturer.
Contact instrument manufacturer.
Contact the Technical Support Dept.
Contact the Technical Support Dept.
Check the position of the union joints, place them in the correct place and make
sure there is enough distilled water.
Use distilled water.
Clean and dry the instruments before sterilising them.
Sterilise the instruments separating them by kind of material.
Do not introduce rusty instruments or instruments that are about to rust because
they affect the other instruments.
Use distilled water (see section 9.2).
Load the printer with paper.
Install a new ink ribbon (see printer manual).
Place the paper correctly.
Connect the printer.
Go to the printer menu (see section 4.3.4) and select a number of copies above 0.
Contact the Technical Support Dept.
Contact the Technical Support Dept.
Clean and/or change the tube.
Connect it properly.
Contact the Technical Support Dept.
Set the time, even if it is correct: change it up or down 1 minute. If the problem
persists, please contact the Technical Support Dept.
Introduce the emptying or filling tube, let all the water out and blow inside
the tube.