Chauvin Arnoux E6N Manual De Instrucciones página 20

Pinza amperimétrica
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

English
TYPICAL NOISE LEVEL AT THE OUTPUT (RMS VALUES)
Frequency band
10 mV/A range
1 V/A range
DISTORTION QUANTITIES
- Position of the conductor in the jaws: 0.5 % of the reading (max) for a conductor moved
from one end of the window to the other.
- Temperature: 800 ppm/°C + 10 mA/°C (max) for a temperature between -10 °C and
+50 °C.
- Load: 0.1% of the reading (max) for a load between 2 kΩ and 10 kΩ on the 100 A/V range
and 2% of the reading (max) for a load of between 10 kΩ and 100 kΩ on the 1 A/V range.
- Frequency: see curves in APPENDIX.
DC – 1 Hz
1 Hz – 10 kHz
15 µV
140 µV
1.5 mV
14 mV
20
10 kHz – 100 kHz
180 µV
18 mV