Descargar Imprimir esta página

Beurer BM 90 Instrucciones De Uso página 99

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 38
Guidance and manufacturer's declaration - electromagnetic immunity
The BM90 is intended for use in the electromagnetic environment specified below. The customer or the user of
BM90 should assure that it is used in such an environment.
IMMuNITY test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge IEC
61000-4-5
Voltage dips, short in-
terruptions and voltage
variations on power
supply input lines IEC
61000-4-11
Power frequency
(50Hz) magnetic
field IEC 61000-4-8
NOTe U
is the a.c. mains voltage prior to application of the test level.
T
For all Me eQuIPMeNT and Me sYsTeMs
IeC 60601 test level
±6 kV contact
±8 kV air
±2 kV for
power supply lines
±1 kV line(s)
to line(s)
<5% U
T
(>95% dip in U
)
T
for 0.5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 sec
3 A/m
Table 2
Compliance level
±6 kV contact
±8 kV air
±2kV for
power supply lines
±1 kV
differential mode
<5% U
T
(>95% dip in U
)
T
for 0.5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 sec
3 A/m
99
electromagnetic environ-
ment - guidance
Floors should be wood, con-
crete or ceramic tile. If floor
are covered with synthetic
material, the relative humidity
should be at least 30%.
Mains power quality should
be that of a typical commer-
cial or hospital environment.
Mains power quality should
be that of a typical commer-
cial or hospital environment.
Mains power quality should
be that of a typical commer-
cial or hospital environment.
If the user of the BM90
requires continued operation
during power mains inter-
ruptions, it is recommended
that the BM90 be powered
from an uninterruptible power
supply or a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical lo-
cation in a typical commercial
or hospital environment.

Publicidad

loading